Elgar CW-M Series
Switch Position Function
ON Required to enable current master and slave operation 1,2
OFF Stand-alone unit or phase slave
ON Current slave 3
OFF Current master
ON Phase slave 4
OFF Phase master
ON Output disabled at power-up, Fault LED lit
Note: If the output switch is in the ON position at power-up,
it has to be moved to the OFF/RESET position, then back
to the ON position to enable the output
5
OFF Output enabled at power-up
6–7 — Reserved for future expansion
ON Constant Current mode 8
OFF Current Shut-off (fault) mode
9–12 — Phase angle selections (see Table 4–2 on page 31)
Table 4–1 CONFIG Dipswitch Functions
4.2 Setting the CONFIG Dipswitch
You may change the CONFIG dipswitch settings to change the default characteristics of
the unit as follows:
• If you would like the power supply to turn its output off, and generate a fault
indication if its output current ever reaches the level that you set from the front
panel, then set switch 8 to the off position.
1 2 3 4 5 6 7 8 9 10 11 12
ON
OFF
• If you would like the power supply to always start up at OV (fault mode) each
time that power is applied, then set switch 5 to the on position.
1 2 3 4 5 6 7 8 9 10 11 12
ON
OFF