Elgar SW5250A Manual de servicio

Busca en linea o descarga Manual de servicio para Amplificadores de audio Elgar SW5250A. Elgar SW5250A Service manual [en] Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 84
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Contact Information
Telephone: 800 733 5427 (toll free in North America)
858 450 0085 (direct)
Fax: 858 458 0267
Email:
Domestic Sales: domorders.sd@ametek.com
International Sales: intl[email protected]
Customer Service: service.ppd@ametek.com
Web: www.programmablepower.com
TrueWave
Switching Amplifier
Service Manual
Models:
TW5250
TW3500
TW1750
This Service Manual is incomplete without the
TrueWave Operation Manual, which contains
detailed descriptions of the TrueWave system,
installation instructions, and operating instructions.
March 2011 Document No. M161469-03 Rev C
Vista de pagina 0
1 2 3 4 5 6 ... 83 84

Indice de contenidos

Pagina 1 - TrueWave

Contact Information Telephone: 800 733 5427 (toll free in North America) 858 450 0085 (direct) Fax: 858 458 0267 Em

Pagina 2

TABLE OF CONTENTS TW SERIES viii This page intentionally left blank.

Pagina 3

1-1 SECTION 1 – THEORY OF OPERATION 1.1 INTRODUCTION This service manual is intended to assist in the maintenance, troubleshooting, repair to the m

Pagina 4

THEORY OF OPERATION TW SERIES 1-2 FIGURE 1-1. TW5250PWR MOD: ØARECTIFIER-OR-PFCDC/DCPWR MOD: ØBRECTIFIER-OR-PFCDC/DCPWR MOD: ØCRECTIFIER-OR-PFCDC/

Pagina 5 - SAFETY SYMBOLS

SERVICE MANUAL THEORY OF OPERATION 1-3 The system shown in Figure 1-1 is a TW5250 meaning that it contains three identical output channels A, B, and

Pagina 6

THEORY OF OPERATION TW SERIES 1-4 1.4 DIGITAL CONTROL BOARD FIGURE 1-2. DIGITAL CONTROL BOARD BLOCK DIAGRAM The Digital Control Board is

Pagina 7 - TABLE OF CONTENTS

SERVICE MANUAL THEORY OF OPERATION 1-5 checked for corruption, and if no corruption is found program control is passed to the FLASH memory. The Digit

Pagina 8 - SECTION 3 – CALIBRATION

THEORY OF OPERATION TW SERIES 1-6 1.5 ANALOG PROCESSOR BOARD FIGURE 1-3. ANALOG PROCESSOR BOARD BLOCK DIAGRAM The Analog Processor Board

Pagina 9 - LIST OF TABLES

SERVICE MANUAL THEORY OF OPERATION 1-7 The sinewave references are fed either directly to the servo amplifiers or are AC coupled, depending on the se

Pagina 10 - TABLE OF CONTENTS TW SERIES

THEORY OF OPERATION TW SERIES 1-8 1.7 POWER CONDITIONER MODULE FIGURE 1-7. POWER CONDITIONER MODULE BLOCK DIAGRAM The power condit

Pagina 11 - 1.2 SYSTEM OVERVIEW

SERVICE MANUAL THEORY OF OPERATION 1-9 The PFC module is a two–board assembly consisting of a PFC Power and a PFC Control board. The single–phase AC

Pagina 13 - 1.3 INTERCONNECTION

THEORY OF OPERATION TW SERIES 1-10 overall switching frequency of 400 kHz. This high switching frequency allows smaller filtering components to be us

Pagina 14 - 1.4 DIGITAL CONTROL BOARD

2-1 SECTION 2 – MAINTENANCE AND TROUBLESHOOTING 2.1 GENERAL This section contains procedures for corrective maintenance of the TrueWave products. In

Pagina 15

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-2 2.3 REQUIRED TEST EQUIPMENT The test equipment required to conduct performance verification procedures

Pagina 16 - 1.5 ANALOG PROCESSOR BOARD

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-3 2.5 FAULT SYMPTOMS/TROUBLESHOOTING The troubleshooting approach taken in this section will be at

Pagina 17

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-4 TABLE 2-3. TROUBLESHOOTING Fault Symptom Troubleshooting/Corrective Action With unit turned on, front

Pagina 18

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-5 Fault Symptom Troubleshooting/Corrective Action Unit comes up normally. Upon relay closure unit

Pagina 19 - CAUTION

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-6 2.6.1 Remove the (20) flat-head screws securing the top cover to the chassis. Note that some of the s

Pagina 20 - 1.9 GLOSSARY

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-7 2.6.3 Remove the fan panel power cables from the Analog board connectors, J10 and J11, as shown

Pagina 21 - 2.2 FACTORY REPAIR

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-8 2.6.5 Disconnect fan panel power cable at J3 on the Housekeeping board mounted on the chassis floor.

Pagina 22 - 2.4 PERIODIC MAINTENANCE

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-9 2.6.7 Gently remove the retaining clip as shown above. The switch can now be slid back through

Pagina 23

i About AMETEK AMETEK Programmable Power, Inc., a Division of AMETEK, Inc., is a global leader in the design and manufacture of precision, programm

Pagina 24

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-10 2.6.9 To remove the Analog and Control boards from the front panel assembly, remove the (8) mountin

Pagina 25

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-11 2.6.11 Remove the (2, 4 or 6) screws, depending on model, that secure the top bracket to the mo

Pagina 26

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-12 2.6.13 Remove the (4) screws that secure the Signal board to the air baffle. 2.6.14 The Signal b

Pagina 27

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-13 2.6.15 Cut all tie-wraps securing the cable assembly to the air baffle as shown above. 2.6

Pagina 28

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-14 2.6.17 Remove the air baffle by lifting straight out as shown above. 2.6.18 Any module can now b

Pagina 29

SERVICE MANUAL MAINTENANCE AND TROUBLESHOOTING 2-15 2.6.19 … And then lifting the module straight up and out of the chassis as shown above. 2.6.

Pagina 30

MAINTENANCE AND TROUBLESHOOTING TW SERIES 2-16 This page intentionally left blank.

Pagina 31

3-1 SECTION 3 – CALIBRATION 3.1 SCOPE These procedures cover the calibration of TrueWave models TW1750, TW3500, and TW5250. Calibrating the TrueWav

Pagina 32

CALIBRATION TW SERIES 3-2 3.4 SETUP In each calibration section the appropriate setup is listed for the calibration step. WARNING If the TrueWave i

Pagina 33

SERVICE MANUAL CALIBRATION 3-3 CAL{1,2,3}:OUTP:REMOTE:HIR:VOLT:OFFSET 0.0 CAL{1,2,3}:OUTP:REM:HIR:VOLT:FREQCAL 1.0 1.0 1.0 1.0 1.0 CAL{1,2,3}:MEAS:LO

Pagina 34

ii This page intentionally left blank.

Pagina 35

CALIBRATION TW SERIES 3-4 Wait for the TW to complete its coupling change to DC. Continue with the following GPIB commands: SOUR:VOLT:RANGE LOW SOUR

Pagina 36

SERVICE MANUAL CALIBRATION 3-5 Send the following query command to the TW and record the return string as MEAS1: MEAS{1,2,3}:VOLT? Calculate the off

Pagina 37 - 3.3 REQUIRED TEST EQUIPMENT

CALIBRATION TW SERIES 3-6 Calculate the gains and offset for the phase being calibrated: GAIN1 = (MEAS1 – MEAS2) / (RDGN1 – RDGN2) GAIN2 = (440.0

Pagina 38 - 3.5 CALIBRATION RESET

SERVICE MANUAL CALIBRATION 3-7 Measure the current offset by sending the following GPIB query and record the return string as MEAS1: MEAS{1,2,3}:CURR

Pagina 39 - 3.6 DC OFFSET CALIBRATION

CALIBRATION TW SERIES 3-8 3.9 DC REMOTE LOW RANGE CALIBRATION Both the measurement and output subsystems are calibrated at the same time. Connect t

Pagina 40

SERVICE MANUAL CALIBRATION 3-9 Wait for the DMM to settle. Take a DMM reading and record the value as RDGN1. Query the output voltage of the TW and

Pagina 41

CALIBRATION TW SERIES 3-10 Wait for the DMM to settle and record the value as RDGN1. Send the following query command to the TW and record the return

Pagina 42

SERVICE MANUAL CALIBRATION 3-11 Calculate the gains and offset for the phase being calibrated: GAIN1 = (MEAS1 – MEAS2) / (RDGN1 – RDGN2) GAIN2 =

Pagina 43

CALIBRATION TW SERIES 3-12 Set the TW to 320 Hz. SOUR:FREQ 320.0 Wait for the DMM to settle and record the voltage as MEAS4. Set the TW to 500 Hz. SO

Pagina 44

SERVICE MANUAL CALIBRATION 3-13 Set the TW’s frequency to 160 Hz with the following GPIB command: SOUR:FREQ 160.0 Wait for the DMM reading to settle

Pagina 45

iii Important Safety Instructions Before applying power to the system, verify that your product is configured properly for your particular applicat

Pagina 46

CALIBRATION TW SERIES 3-14 3.12 LOCAL HIGH RANGE FREQUENCY CALIBRATION Connect the TW output to the DMM. Set the DMM up for auto–ranging AC voltage.

Pagina 47

SERVICE MANUAL CALIBRATION 3-15 Calculate the calibration constants: VALUE1 = 240.0 / MEAS1 VALUE2 = 240.0 / MEAS2 VALUE3 = 240.0 / MEAS3 VALUE4 =

Pagina 48

CALIBRATION TW SERIES 3-16 Wait for the DMM reading to settle and record the value as RDNG4. Query the TW’s output put voltage with the GPIB string b

Pagina 49

SERVICE MANUAL CALIBRATION 3-17 Close the TW’s output relay: OUTPUT ON Wait for the DMM to settle and record the voltage as MEAS1. Set the TW to 80 H

Pagina 50

CALIBRATION TW SERIES 3-18 Close the TW’s output relay: OUTPUT ON Wait for the DMM reading to settle and record the value as RDNG1. Query the TW’s ou

Pagina 51

SERVICE MANUAL CALIBRATION 3-19 Query the TW’s output put voltage with the GPIB string below and record this value as MEAS5. MEAS{1,2,3}:VOLT? Set th

Pagina 52

CALIBRATION TW SERIES 3-20 Set the TW to 160 Hz. SOUR:FREQ 160.0 Wait for the DMM to settle and record the voltage as MEAS3. Set the TW to 320 Hz. SO

Pagina 53

SERVICE MANUAL CALIBRATION 3-21 Set the TW’s frequency to 80 Hz with the following GPIB command: SOUR:FREQ 80.0 Wait for the DMM reading to settle an

Pagina 54

CALIBRATION TW SERIES 3-22 Calculate the measurement system frequency compensation values: VALUE1 = RDGN1 / MEAS1 VALUE2 = RDGN2 / MEAS2 VALUE3 = R

Pagina 55

SERVICE MANUAL CALIBRATION 3-23 Wait for the phase angle measurement to stabilize on the counter timer. Record the phase angle as RDGN4. Set the TW

Pagina 56

iv Product Family: Models: TW5250, TW3500, TW1750 Warranty Period: One Year WARRANTY TERMS AMETEK Programmable Power, Inc. (“AMETEK”), provides this

Pagina 57

CALIBRATION TW SERIES 3-24 Query the TW for the phase angle with the following command and record the value as MEAS1. MEAS2:PHASE? Set the TW frequen

Pagina 58

SERVICE MANUAL CALIBRATION 3-25 Set the TW frequency to 60 Hz and open the output via the GPIB: SOUR:FREQ 60.0 OUTP OFF Calculate the phase offset ca

Pagina 59

CALIBRATION TW SERIES 3-26 Wait for the counter timer to settle and record the phase angle as RDNG1. Query the TW for the phase angle with the follow

Pagina 60

SERVICE MANUAL CALIBRATION 3-27 Set the TW frequency to 60 Hz and open the output via the GPIB: SOUR:FREQ 60.0 OUTP OFF Calculate the phase offset ca

Pagina 61

CALIBRATION TW SERIES 3-28 Wait for the DMM to settle and record the value as RDNG2. Calculate the gain and offset for the external input: GAIN = 6

Pagina 62

SERVICE MANUAL CALIBRATION 3-29 Reset the TW back to AC coupling mode, output off. SOUR{1,2,3}:VOLT:OFFSET 0.0 OUTP OFF OUTP:COUP AC Calculate th

Pagina 63

CALIBRATION TW SERIES 3-30 Query the TW’s watts using the GPIB string below, and record this value as MEAS2. MEAS{1,2,3}:POW? Turn the TW’s output o

Pagina 64

SERVICE MANUAL CALIBRATION 3-31 Turn the TW’s output off via the GPIB: OUTP OFF Calculate the gain and offset values for updating the TW: GAIN =

Pagina 65

CALIBRATION TW SERIES 3-32 Program the TW to 160 Hz using the string below: SOUR{1,2,3}:FREQ 160 Wait until the power analyzer has stabilized and rec

Pagina 66

SERVICE MANUAL CALIBRATION 3-33 3.23 LOCAL LOW RANGE WATTS FREQUENCY CALIBRATION Connect the 12 ohm resistor to the output of the TW through the powe

Pagina 67

v TABLE OF CONTENTS Warranty...

Pagina 68

CALIBRATION TW SERIES 3-34 Query the TW’s watts using the GPIB string below, and record this value as MEAS4. MEAS{1,2,3}:POW? Program the TW to 500 H

Pagina 69

SERVICE MANUAL CALIBRATION 3-35 Query the TW’s watts using the GPIB string below, and record this value as MEAS1. MEAS{1,2,3}:POW? Program the TW to

Pagina 70

CALIBRATION TW SERIES 3-36 Calculate the calibration constants: VALUE1 = RDGN1/MEAS1 VALUE2 = RDGN2/MEAS2 VALUE3 = RDGN3/MEAS3 VALUE4 = RDGN4

Pagina 71

SERVICE MANUAL CALIBRATION 3-37 Calculate the gain from the collected data: GAIN = (RDGN2 – RDNG1) / (MEAS2 – MEAS1) Update the TW’s gain setting u

Pagina 72

CALIBRATION TW SERIES 3-38 Update the TW’s gain and offset calibration values with the following GPIB strings: CAL{1,2,3}:MEAS:LOC:HIR:WATT:GAIN GAIN

Pagina 73

SERVICE MANUAL CALIBRATION 3-39 3.28 HIGH RANGE CURRENT FREQUENCY CALIBRATION Connect the 48 ohm resistor to the output of the TW through the power a

Pagina 74

CALIBRATION TW SERIES 3-40 Query the TW’s current using the GPIB string below, and record this value as MEAS4. MEAS{1,2,3}:CURR? Program the TW to 50

Pagina 75

SERVICE MANUAL CALIBRATION 3-41 Wait until the power analyzer has stabilized and record the watts reading as RDNG1. Query the TW’s watts using the GP

Pagina 76

CALIBRATION TW SERIES 3-42 Calculate the calibration constants: VALUE1 = RDGN1/MEAS1 VALUE2 = RDGN2/MEAS2 VALUE3 = RDGN3/MEAS3 VALUE4 = RDGN4

Pagina 77

SERVICE MANUAL CALIBRATION 3-43 Wait until the power analyzer has stabilized and record the watts reading as RDNG3. Query the TW’s watts using the GP

Pagina 78

TABLE OF CONTENTS TW SERIES viSECTION 3 – CALIBRATION 3.1 SCOPE...

Pagina 79

CALIBRATION TW SERIES 3-44 This page intentionally left blank.

Pagina 80 - CALIBRATION TW SERIES

4-1 SECTION 4 – PARTS LIST 4.1 GENERAL This section contains the top assembly parts list for the TrueWave products. The parts lists below directly c

Pagina 81 - 4.2 PARTS LIST

PARTS LIST TW SERIES 4-2 4.3 ORDERING SPARE PARTS Contact Elgar Electronics Corporation to order spare parts or assemblies. Please specify the asse

Pagina 82 - 4.3 ORDERING SPARE PARTS

5-1 SECTION 5 – DIAGRAMS 5.1 GENERAL This section contains the interconnect diagrams and top assembly diagrams for the TrueWave series. The intercon

Pagina 83 - 5.2 DIAGRAMS

DIAGRAMS TW SERIES 5-2 This page intentionally left blank.

Pagina 84 - DIAGRAMS TW SERIES

SERVICE MANUAL TABLE OF CONTENTS viiSECTION 4 – PARTS LIST 4.1 GENERAL ...................

Comentarios a estos manuales

Sin comentarios